<4. OPERATION>
4-8
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
<Time chart>
Example: N (number of cleaning repetitions) = 4
SV1
(Draining)
SV3
(Water sample)
SV4
(Zero water)
OFF
OFF
OFF
OFF
ON
ON
OFF
ON
N*2 = 1
2
3
4
ON
HOLD ON
Cleaning start
Cleaning time 2
One cleaning cycle*3
Recovery time
Drain time 2
Water fill time 2
Water fill time 2
SV2
(Cleaning)
Output
holding *1
F0408.ai
*1: During cleaning, the output is always in the hold state.
The display is also set to be held using FUNCTION "8." in the <PROGRAM1> mode.
Contact outputs C1 and C2 can be used as contacts for the period during cleaning and
calibration. This is set using FUNCTION "E." in the <PROGRAM1> mode.
*2: N; Number of cleaning repetitions
*3: One automatic cleaning cycle = N (dr f cleaning) + (dr f recovery)
Figure 4.8 Automatic-cleaning Operation Time Cheat (Suffix code: -A3)
Table 4.3 Setpoints for Automatic Cleaning and Automatic Zero Calibration
Item
Setting
range
Factory
setting
Unit
Setting mode
Mode
FUNCTION
Cleaning period
0.1 to 24.0
2.0
Hours
PROGRAM 2
1
Cleaning calibration ratio
1 to 10
1
times
PROGRAM 2
2
Cleaning time 2
10 to 120
30
Sec
PROGRAM 2
5
Draining time 2
10 to 120
10
Sec
PROGRAM 2
7
Water filling time 2
10 to 120
100
Sec
PROGRAM 2
9
Relaxation time
30 to 600
150
Sec
PROGRAM 2
A
Number of cleaning repetitions
1 to 20
5
times
PROGRAM 2
3
For the setting procedures for each setpoint, see Section 6.7.