![YOKOGAWA EXA TB TB400G Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/yokogawa/exa-tb-tb400g/exa-tb-tb400g_user-manual_911250071.webp)
<4. OPERATION>
4-6
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
<Time chart>
Example: N (number of cleaning repetitions) = 4
F0406.ai
*1 During cleaning, the output is always in the hold state.
The display is also set to be held using FUNCTION "8." in the <PROGRAM1> mode.
Contact outputs C1 and C2 can be used as contacts for the period during cleaning and
calibration. This is set using FUNCTION "E." in the <PROGRAM1> mode.
*2: N; Number of cleaning repeats
*3: One automatic cleaning cycle = N (cl dr water filling) + relaxation
SV1
(Drain)
OFF
OFF
N*2 =
ON
1
2
3
4
ON
OFF
HOLD
HOLD ON
Power on
Measurement
start
Cleaning cycle
Cleaning time 1
One cleaning cycle*3
Recovery time
Drain time 1
Water fill time 1
SV2
(Cleaning)
Output
holding *1
Figure 4.6 Automatic-cleaning Operation Time Chart (Suffix code: -A2)
Table 4.2 Cleaning Operation Setpoints
Item
Setting
range
Factory
setting
Unit
Setting mode
Mode
FUNCTION
Cleaning period
0.1 to 24.0
2.0
Hours
PROGRAM 2
1
Cleaning time 1
10 to 120
30
Sec
PROGRAM 2
4
Draining time 1
10 to 120
10
Sec
PROGRAM 2
6
Water filling time 1
10 to 120
100
Sec
PROGRAM 2
8
Relaxation time
30 to 600
150
Sec
PROGRAM 2
A
Number of cleaning repetitions
1 to 20
5
times
PROGRAM 2
3
See Section 6.7 for the setting procedures for each setpoint.