<6. PROCEDURES FOR CONVERTER OPERATION>
6-34
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
(27) Selecting AUX (C1, C2) Contact Output (FUNCTION "E.")
The operating conditions for AUX contact output (C1, C2) are selected from the following two.
•
High and low alarm contact output (setting: on)
• Contact output during automatic cleaning and automatic zero-calibration operations (setting:
off)
The following shows the setting key operations.
(Example of display)
(1) Select FUNCTION "E." in the <PROGRAM1>
=> ON
mode. The current set status is displayed.
(2) Press the [>] key to select on/off.
=> OFF
F6601.ai
ON
OFF
(3) Press the [ENT] key to enter the setting.
=> OFF
The display flashes once.
(28) Selecting Execution/Stop of Spike Detection Function (FUNCTION "F.")
(Example of display)
(1) Select FUNCTION "F." in the <PROGRAM1>
=> OFF (stop)
mode. The current set status is displayed.
(2) Press the [>] key to select on/off.
=> ON (execution)
F6601.ai
ON
OFF
(3) Press the [ENT] key to enter the setting.
=> ON
The display flashes once.