<6. PROCEDURES FOR CONVERTER OPERATION>
6-21
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
(1) Setting Output Ranges 1, 2, and 3 (Functions "1", "2", and "3")
The analog output is sent out by being selected for one of the output ranges, 1, 2, or 3.
These are three output range selections as shown below.
• Auto-range selection
• Remote range selection
• Local range selection
For details on selection, see (11), (12), and (13) in Section 6.6. In this item, the setting of upper
and lower limit values in output ranges 1, 2, and 3 is described taking output range 1 as an
example.
(Example of display)
(1) Select FUNCTION "1" in the <PROGRAM1>
=> L00.00 mg/L
mode. If the lower limit is not to be changed,
start operation from step 6.
(2
) Press the [>] key.
=> L0000. mg/L
This display is now ready for
determination of the decimal point
position.
(3
)
Determine the decimal point position with the
=> L00.00 mg/L
[>] key.
Every time the [>] key is pressed, the decimal
point moves to the right one digit and when the
least significant digit is reached, it returns to the
most significant digit.
(4) Press the [ENT] key.
=> L
0
0.00 mg/L
The most significant digit flashes and
wits for the numerical value entry.
(5)
Select a digit with the [>] key and set the
=> L0
5
.00 mg/L
desired lower limit using the [^] or [v] key.
Every time the [>] key is pressed, the flashing
digit moves to the right one digit and when the
least significant digit is reached, it returns to the
most significant digit.
(6) When press [ENT] key, current upper limit is
=> H10.00 mg/L
indicated.
(7) Press the [>] key.
=> H0000. mg/L
The display is now ready for
determination of the decimal point
position.
(8)
Determine the position of the decimal point
=> H000.0 mg/L
using the [>] key.
Every time the [>] key is pressed, the decimal
point moves to the right one digit and when it
reaches the least significant digit, it returns to
the most significant digit.
(9) Press the [ENT] key.
=> H
0
00.0 mg/L
The most significant digit flashes and
waits for the numerical value entry.
(10) Select a digit and set the desired upper limit
=> H0
2
0.0 mg/L
using the [^] or [v] key.
Every time the [>] key is pressed, the decimal
point moves to the right one digit and when it
reaches the least significant digit, it returns to
the most significant digit.
Содержание EXA TB TB400G
Страница 65: ...Blank Page ...
Страница 87: ...Blank Page ...
Страница 125: ...Blank Page ...
Страница 143: ...Blank Page ...
Страница 149: ...Blank Page ...
Страница 151: ...Blank Page ...
Страница 157: ...Blank Page ...