<6. PROCEDURES FOR CONVERTER OPERATION>
6-33
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
(25) Selecting Designation of Units (FUNCTION "C.")
This function can set the unit of turbidity to be displayed on the converter at either "mg/L",
"degree", or "FTU".
If the unit designation is to be changes, perform it using the following key operations.
(Example of display)
(1) Select FUNCTION "C." in the <PROGRAM1>
=> _ _1 (mg/L)
mode. The current set code is displayed.
(2) Press the [>] key to select the setting code.
=> _ _2 (Degree)
F6605.ai
_ _1
_ _2
(mg/l)
(Degree)
_ _3
(FTU)
(3) Press the [ENT] key to enter the setting.
=> _ _2
The display flashes once.
The unit selected in step 3 is lit in the unit display area on the right of the data display when the
turbidity is displayed (Figure 6.4).
MEAS.
MAINT.
PROGRAM 1
PROGRAM 2
mg/l
FTU
nA mA
V %
}
Unit set in
FUNCTION "C."
is lit.
Turbidity reading
F0604.ai
Figure 6.4 Unit Designation
(26) Selecting Calibration Plate/Standard solution (FUNCTION "D.")
In the span calibration operation for FUNCTION "4" in the <MAINT.> mode, this function sets
either the calibration plate or the standard solution for span calibration.
Upon shipment from the factory, on (calibration plate) is set. If off (standard solution) is to be
selected, perform the key operations in the following procedure.
(Example of display)
(1) Select FUNCTION "D." in the <PROGRAM1>
=> ON (calibration plate)
mode. The current set status is displayed.
(2) Press the [>] key to select on/off.
=> OFF (standard solution)
F6601.ai
ON
OFF
(3) Press the [ENT] key to enter the setting.
=> OFF
The display flashes once.