<6. PROCEDURES FOR CONVERTER OPERATION>
6-25
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
(7) Setting Low Alarm Value (FUNCTION "9")
In the high and low alarm detection function is set to "execution" (FUNCTION "9." in
<PROGRAM2> mode), when a measured value is below this setpoint, the MEAS lamp flashes
and if AUX (C1, C2) contacts are set for high and low alarms (FUNCTION "E." in <PROGRAM1>
mode), the contact output is actuated. Alarm hysteresis is 2% of the setpoint.
The following shows the setting key operations.
(Example of display)
(1) Select FUNCTION "9" in the <PROGRAM1>
=> -010.0 mg/L
mode.
(2) Press the [>] key.
=>
-
010.0 mg/L
The most significant digit flashes.
(3) Select a digit with the [>] key and set it to the
=> 00
0
0.0 mg/L
desired value using the [^] or [v] key.
The flashing digit moves to the right
and when the least significant digit is
reached, it returns to the most
significant digit.
(4) Press the [ENT] key to enter the set value.
=> 0000.0 mg/L
The entire display flashes once.
In step 4, if the set value is beyond the permissible range (-10 to 2200 mg/L), "not" is displayed.
Reset the message by pressing the [>] key and repeat the operations from step 2.
(8) Setting Turbidity Value of Calibration Plate (FUNCTION "A")
Upon shipment from the factory, the value of the attached calibration plate has been set.
The following shows the setting key operations.
(Example of display)
(1) Select FUNCTION "A" in the <PROGRAM1>
=> 0090.0 mg/L
mode.
(2) Press the [>] key.
=>
0
090.0 mg/L
The most significant digit flashes.
(3) Select a digit with the [>] key and set it to the
=> 009
5
.0 mg/L
desired value using the [^] or [v] key.
The flashing digit moves to the right
and when the least significant digit is
reached, it returns to the most
significant digit.
(4) Press the [ENT] key to enter the set value.
=> 0095.0 mg/L
The entire display flashes once.
In step 4, if the set value is beyond the permissible range (0 to 2000 mg/L), "not" is displayed.
Reset the message by pressing the [>] key and repeat the operations from step 2.