<2. SPECIFICATIONS>
2-38
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
2.7 Internal Wiring Diagram
Figure 2.23 shows the internal wiring diagram of the turbidimeter with a sampling system. For
external wiring, see Section 3.3.
If the detector and the converter are to be delivered as separate units, a dedicated cable is
shipped connected to the detector.
V1
V2
SV2
V3
SV1
SV3
V4
SV4
V5
V6
V7
V8
V9
V10
0
S
C
1
2
1
2
S1
S2
V+
V-
AG
P1
P2
S1
S2
V+
V-
AG
P1
P2
G L1 L2
Turbidity
element
Reference
element
Lamp
Detector
Dedicated cable
Converter
Valves*1
*1 : Valve connections vary with the specifications as shown Table 2.2.
F0223.ai
Figure 2.23 Internal Wiring Diagram
Table 2.2
Specifications and Connections
Suffix code Specifications with/without sampling system, automatic cleaning
and zero calibration
Valves to be
connected
-NN, -KC
W/o sampling system, automatic cleaning, or automatic zero calibration
No valve connection
-A1
W/ sampling system, but w/o automatic cleaning or automatic zero
calibration
No valve connection
-A2
W/ sampling system and automatic cleaning but w/o automatic zero
calibration
SV1, SV2
-A3
W/ sampling system, automatic cleaning, and automatic zero calibration SV1, SV2, SV3, SV4
Содержание EXA TB TB400G