App-9
IM 710105-03E
Appendix
1
2
3
4
5
App
Index
When the Record Length Is 125 M points
(This record length can be selected on DLM2000s with the /M2 or /M3 option)
Intpl: Interpolation Mode
Rep: Repetitive Sampling Mode
Settings
Time/div
When High Resolution mode is off
When High Resolution mode is on
When Interleave mode is off*
When Interleave mode is on
When Interleave mode is off
When Interleave mode is on
Realtime
Intpl/Rep
Realtime
Intpl/Rep
Realtime
Intpl/Rep
Realtime
Intpl/Rep
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
500 s
25k
125M
Interpolation
mode and
repetitive
sampling mode
cannot be
used when you
specify this
record length.
25 k
125 M
Interpolation
mode and
repetitive
sampling mode
cannot be
used when you
specify this
record length.
When you specify this record length, high resolution mode cannot be
turned on.
200 s
62.5k
125M
62.5 k 125 M
100 s
125k
125M
125 k
125 M
50 s
250k
125M
250 k
125 M
20 s
625k
125M
625 k
125 M
10 s
1.25M 125M
1.25 M 125 M
5 s
2.5M
125M
2.5 M
125 M
2 s
6.25M 125M
6.25 M 125 M
1 s
12.5M 125M
12.5 M 125 M
500 ms 25M
125M
25 M
125 M
200 ms 62.5M 125M
62.5 M 125 M
100 ms 125M
125M
125 M 125 M
50 ms 250M
125M
250 M 125 M
20 ms 625M
125M
625 M 125 M
10 ms 1.25G 125M
1.25 G 125 M
5 ms
1.25G 62.5M
2.5 G
125 M
2 ms
1.25G 25M
2.5 G
50 M
1 ms
1.25G 12.5M
2.5 G
25 M
500 μs
1.25G 6.25M
2.5 G
12.5 M
200 μs
1.25G 2.5M
2.5 G
5 M
100 μs
1.25G 1.25M
2.5 G
2.5 M
50 μs
1.25G 625k
2.5 G
1.25 M
20 μs
1.25G 250k
2.5 G
500 k
10 μs
1.25G 125k
2.5 G
250 k
5 μs
1.25G 62.5k
2.5 G
125 k
2 μs
1.25G 25k
2.5 G
50 k
1 μs
1.25G 12.5k
2.5 G
25 k
500 ns 1.25G 6.25k
2.5 G
12.5 k
200 ns 1.25G 2.5k
2.5 G
5 k
100 ns 1.25G 1.25k
2.5 G
2.5 k
50 ns
1.25G 625
2.5 G
1.25 k
20 ns
1.25G 250
2.5 G
500
10 ns
1.25G 125
2.5 G
250
5 ns
1.25G 62.5
2.5 G
125
2 ns
1.25G 25
2.5 G
50
1 ns
1.25G 12.5
2.5 G
25
Acquisition when you specify this record length is performed in single mode.
* On DLM2000s with the /M2 option, with this record length, interleave mode cannot be turned off.
: Roll Mode
Appendix 1 Relationship between the Time Axis Setting, Record Length, and Sample Rate