App-8
IM 710105-03E
When the Record Length Is 62.5 M points
(This record length can be selected on DLM2000s with the /M1(S) or /M2 option)
Intpl: Interpolation Mode
Rep: Repetitive Sampling Mode
Settings
Time/div
When High Resolution mode is off
When High Resolution mode is on
When Interleave mode is off*
When Interleave mode is on
When Interleave mode is off
When Interleave mode is on
Realtime
Intpl/Rep
Realtime
Intpl/Rep
Realtime
Intpl/Rep
Realtime
Intpl/Rep
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
Sample
Rate
(S/s)
Record
Length
(Points)
500 s
12.5 k 62.5 M
Interpolation
mode and
repetitive
sampling mode
cannot be
used when you
specify this
record length.
12.5 k 62.5 M
Interpolation
mode and
repetitive
sampling mode
cannot be
used when you
specify this
record length.
When you specify this record length, high resolution mode cannot be
turned on.
200 s
31.25 k 62.5 M
31.25 k 62.5 M
100 s
62.5 k 62.5 M
62.5 k 62.5 M
50 s
125 k
62.5 M
125 k
62.5 M
20 s
312.5 k 62.5 M
312.5 k 62.5 M
10 s
625 k
62.5 M
625 k
62.5 M
5 s
1.25 M 62.5 M
1.25 M 62.5 M
2 s
3.125 M 62.5 M
3.125 M 62.5 M
1 s
6.25 M 62.5 M
6.25 M 62.5 M
500 ms 12.5 M 62.5 M
12.5 M 62.5 M
200 ms 31.25 M 62.5 M
31.25 M 62.5 M
100 ms 62.5 M 62.5 M
62.5 M 62.5 M
50 ms 125 M 62.5 M
125 M 62.5 M
20 ms 312.5 M 62.5 M
312.5 M 62.5 M
10 ms 625 M 62.5 M
625 M 62.5 M
5 ms
1.25 G 62.5 M
1.25 G 62.5 M
2 ms
1.25 G 25 M
2.5 G
50 M
1 ms
1.25 G 12.5 M
2.5 G
25 M
500 μs
1.25 G 6.25 M
2.5 G
12.5 M
200 μs
1.25 G 2.5 M
2.5 G
5 M
100 μs
1.25 G 1.25 M
2.5 G
2.5 M
50 μs
1.25 G 625 k
2.5 G
1.25 M
20 μs
1.25 G 250 k
2.5 G
500 k
10 μs
1.25 G 125 k
2.5 G
250 k
5 μs
1.25 G 62.5 k
2.5 G
125 k
2 μs
1.25 G 25 k
2.5 G
50 k
1 μs
1.25 G 12.5 k
2.5 G
25 k
500 ns 1.25 G 6.25 k
2.5 G
12.5 k
200 ns 1.25 G 2.5 k
2.5 G
5 k
100 ns 1.25 G 1.25 k
2.5 G
2.5 k
50 ns
1.25 G 625
2.5 G
1.25 k
20 ns
1.25 G 250
2.5 G
500
10 ns
1.25 G 125
2.5 G
250
5 ns
1.25 G 62.5
2.5 G
125
2 ns
1.25 G 25
2.5 G
50
1 ns
1.25 G 12.5
2.5 G
25
Acquisition when you specify this record length is performed in single mode.
* On DLM2000s with the /M1(S) option, with this record length, interleave mode cannot be turned off.
: Roll Mode