5-108
IM DLM6054-17EN
:MATH<x>:IPOint:POSition
Function Sets the computation reference point or queries the
current setting.
Syntax
:MATH<x>:IPOint:POSition {<NRf>}
:MATH<x>:IPOint:POSition?
<x> = 1 to 4
<NRf> = –5 to 5 (div)
Example
:MATH1:IPOINT:POSITION 0
:MATH1:IPOINT:POSITION? -> :MATH1:
IPOINT:POSITION 0.000E+00
Description This command is valid when
:MATH<x>:OPERation
INTegral
.
:MATH<x>:LABel?
Function Queries all settings related to the label of the
computed waveform.
Syntax
:MATH<x>:LABel?
<x> = 1 to 4
:MATH<x>:LABel[:DEFine]
Function Sets the label of the computed waveform or queries
the current setting.
Syntax
:MATH<x>:LABel[:DEFine] {<String>}
:MATH<x>:LABel[:DEFine]?
<x> = 1 to 4
<String> = Up to 8 characters
Example
:MATH1:LABEL:DEFINE "MATH1"
:MATH1:LABEL:DEFINE? -> :MATH1:LABEL:
DEFINE "MATH1"
:MATH<x>:LABel:MODE
Function Turns ON/OFF the label display of the computed
waveform or queries the current setting.
Syntax
:MATH<x>:LABel:MODE {<Boolean>}
:MATH<x>:LABel:MODE?
<x> = 1 to 4
:MATH<x>:OPERation
Function Sets the operator or queries the current setting.
Syntax
:MATH<x>:OPERation {DA,GROup<y>|ECOunt,
<NRf>|FILTer,<NRf>|INTegral,<NRf>|
MINus,<NRf>,<NRf>|MULTiple,<NRf>,<NRf>|
PLUS,<NRf>,<NRf>|RCOunt,<NRf>,NRf>|
USERdefine}
<x> = 1 to 4
<y> = 1 to 5
<NRf> = 1 to 5 (you can only specify 5 for MATH2 to 4)
Example
:MATH1:OPERATION FILTER,1
:MATH1:OPERATION? -> :MATH1:
OPERATION FILTER,1
Description For unary operators (ECOunt|FILTer|INTegral), select
the target waveform using the first <NRf>.
For binary operators (MINus|MULTiple|PLUS|RCOu
nt), select the target waveform of the first term using
the first <NRf> and the target waveform of the second
term using the second <NRf>.
On the DLM6000, when the Operation setting is set
to DA, select GROup<x>.
<NRf> is not required for the USERdefine operator.
:MATH<x>:SCALe?
Function Queries all settings related to scaling.
Syntax
:MATH<x>:SCALe?
<x> = 1 to 4
:MATH<x>:SCALe:CENTer
Function Sets the offset of the computed waveform or queries
the current setting.
Syntax
:MATH<x>:SCALe:CENTer {<NRf>|<Voltage>|
<Current>}
:MATH<x>:SCALe:CENTer?
<x> = 1 to 4
<NRf>, <Voltage>, and <Current> = See the DL6000/
DLM6000 User’s Manual.
Example
:MATH1:SCALE:CENTER 1
:MATH1:SCALE:CENTER? -> :MATH1:SCALE:
CENTER 1.000E+00
:MATH<x>:SCALe:MODE
Function Sets the scaling mode or queries the current setting.
Syntax
:MATH<x>:SCALe:MODE {AUTO|MANual}
:MATH<x>:SCALe:MODE?
<x> = 1 to 4
Example
:MATH1:SCALE:MODE AUTO
:MATH1:SCALE:MODE? -> :MATH1:SCALE:
MODE AUTO
5.18 MATH Group