5-61
IM DLM6054-17EN
Commands
1
2
3
4
5
6
App
Index
:CHANnel<x>:PROBe?
Function Queries all settings related to the probe attenuation
of the channel.
Syntax
:CHANneL<x>:PROBe?
<x> = 1 to 4
Example
:CHANNEL1:PROBE? -> :CHANNEL1:PROBE:
MODE 1
:CHANnel<x>:PROBe[:MODE]
Function Sets the probe attenuation of the channel or queries
the current setting.
Syntax
:CHANneL<x>:PROBe[:MODE] {<NRf>|AUTO|
C1|C10|C100}
:CHANneL<x>:PROBe[:MODE]?
<x> = 1 to 4
<NRf> = 1,2,5,10,20,50,100,200,500,1000
Example
:CHANNEL1:PROBE:MODE 1
:CHANNEL1:PROBE:MODE? -> :CHANNEL1:
PROBE:MODE 1
:Channal<x>:PROBe:AUTO?
Function Queries the probe attenuation of the channel when
set to AUTO.
Syntax
:Channal<x>:PROBe:AUTO?
<x> = 1 to 4
Example
:CHANNEL1:PROBE:AUTO? -> :CHANNEL1:
PROBE:AUTO 1
:CHANnel<x>:PZCalibrate[:EXECute]
Function Execute the zero adjustment of the current probe of
the channel.
Syntax
:CHANnel<x>:PZCalibrate[:EXECute]
<x> = 1 to 4
Example
:CHANNEL1:PZCALIBRATE:EXECUTE
:CHANnel<x>:VDIV
Function Sets the vertical sensitivity (V/div) of the channel or
queries the current setting.
Syntax
:CHANnel<x>:VDIV {<Voltage>|<Current>}
:CHANnel<x>:VDIV?
<x> = 1 to 4
<Voltage> and <Current> = See the DL6000/
DLM6000 User’s Manual.
Example
:CHANNEL1:VDIV 5V
:CHANNEL1:VDIV? -> :CHANNEL1:
VDIV 5.000E+00
:CHANnel<x>:LSCale:BVALue
Function Sets or queries an offset B value.
Syntax
:CHANnel<x>:LSCale:BVALue {<NRf>}
:CHANnel<x>:LSCale:BVALue?
<x> = 1 to 4
<NRf> = –1.0000E+31 to 1.0000E+31
Example
:CHANNEL1:LSCALE:BVALUE 10
:CHANNEL1:LSCALE:BVALUE? -> :CHANNEL1:
LSCALE:BVALUE 10.0000E+00
:CHANnel<x>:LSCale:MODE
Function Sets or queries the on/off status of linear scaling.
Syntax
:CHANnel<x>:LSCale:MODE {<Boolean>}
:CHANnel<x>:LSCale:MODE?
<x> = 1 to 4
Example
:CHANNEL1:LSCALE:MODE ON
:CHANNEL1:LSCALE:MODE? -> :CHANNEL1:
LSCALE:MODE 1
:CHANnel<x>:LSCale:UNIT
Function Sets or queries a unit that is attached to linear scaling
results.
Syntax
:CHANnel<x>:LSCale:UNIT {<String>}
:CHANnel<x>:LSCale:UNIT?
<x> = 1 to 4
<String> = Up to four characters
Example
:CHANNEL1:LSCALE:UNIT “ EU”
:CHANNEL1:LSCALE:UNIT? -> :CHANNEL1:
LSCALE:UNIT “ EU”
Description You can only use the characters and symbols that
appear on the DL6000/DLM6000 soft keyboard.
Assignment of a unit never affects the scale value.
:CHANnel<x>:OFFSet
Function Sets the offset voltage of the channel or queries the
current setting.
Syntax
:CHANneL<x>:OFFSet
{<Voltage>|<Current>}
:CHANneL<x>:OFFSet?
<x> = 1 to 4
<Voltage> and <Current> = See the DL6000/
DLM6000 User’s Manual.
Example
:CHANNEL1:OFFSET 1V
:CHANNEL1:OFFSET? -> :CHANNEL1:
OFFSET 1.000E+00
:CHANnel<x>:POSition
Function Sets the vertical position of the channel or queries the
current setting.
Syntax
:CHANneL<x>:POSition {<NRf>}
:CHANneL<x>:POSition?
<x> = 1 to 4
<NRf> = –4 to 4 (div)
Example
:CHANNEL1:POSITION 1
:CHANNEL1:POSITION? -> :CHANNEL1:
POSITION 1.000E+00
5.6 CHANnel Group