5-114
IM DLM6054-17EN
:MEASure:CALCulation:VALue<x>?
Function Queries the automated measured value of the
calculation item.
Syntax
:MEASure:CALCulation:VALue<x>? {<NRf>}
<x> = 1 to 4
<NRf> = 1 to 100000
Example
:MEASURE:CALCULATION:VALUE1?
-> :MEASURE:CALCULATION:VALUE1 1.000E+00
Description • If the measurement is not possible, "NAN (Not A
Number" is returned.
• <NRf> indicates the nth automated measured
value in the past.
In the case of cycle statistical processing, specify
the <NRf>th cycle from the left of the screen.
To specify the oldest automated measured value,
specify 1.
If <NRf> is omitted, the latest automated measured
value is specified.
If the value corresponding to the relevant count is
not present, "NAN (Not A Number)" is returned.
:MEASure:CONTinuous?
Function Queries all settings related to the continuous
statistical processing.
Syntax
:MEASure:CONTinuous?
:MEASure:CONTinuous:COUNt
Function Sets the continuous statistical processing count or
queries the current setting.
Syntax
:MEASure:CONTinuous:COUNt {<NRf>}
:MEASure:CONTinuous:COUNt?
<NRf> = 0 to 100000
Example
:MEASURE:CONTINUOUS:COUNT 10
:MEASURE:CONTINUOUS:COUNT? -> :MEASURE:
CONTINUOUS:COUNT 10
Description When <NRf> = 0, the maximum count that is possible
under the current settings is automatically set.
:MEASure:CONTinuous:RESTart
Function Restarts the continuous statistical processing.
Syntax
:MEASure:CONTinuous:RESTart
Example
:MEASURE:CONTINUOUS:RESTART
Description Clears the previous statistical data.
:MEASure:CYCLe?
Function Queries all settings related to the cycle statistical
processing.
Syntax
:MEASure:CYCLe?
:MEASure:CYCLe:ABORt
Function Aborts the execution of the cycle statistical
processing.
Syntax
:MEASure:CYCLe:ABORt
Example
:MEASURE:CYCLE:ABORT
:MEASure:CYCLe:EXECute
Function Executes the cycle statistical processing. This is an
overlap command.
Syntax
:MEASure:CYCLe:EXECute
Example
:MEASURE:CYCLE:EXECUTE
Description Continues the operation without clearing the previous
statistical data.
:MEASure:CYCLe:TRACe
Function Sets the cycle source trace of the continuous
statistical processing count or queries the current
setting.
Syntax
:MEASure:CYCLe:TRACe {<NRf>|A<x>|B<x>|
C<x>|D<x>}
:MEASure:CYCLe:TRACe?
<NRf> = 1 to 8
<x> = 1 to 7
Example
:MEASURE:CYCLE:TRACE 1
:MEASURE:CYCLE:TRACE? -> :MEASURE:
CYCLE:TRACE 1
Description On 16-bit models, you can only select
{<NRf>|A<y>|C<y>} .
:MEASure:DISPlay
Function Turns ON/OFF the display of the automated
measurement of waveform parameters or queries the
current setting.
Syntax
:MEASure:DISPlay {<Boolean>}
:MEASure:DISPlay?
Example
:MEASURE:DISPLAY ON
:MEASURE:DISPLAY? -> :MEASURE:DISPLAY 1
:MEASure:HISTory:ABORt
Function Aborts the execution of the statistical processing of
the history data.
Syntax
:MEASure:HISTory:ABORt
Example
:MEASURE:HISTORY:ABORT
:MEASure:HISTory:EXECute
Function Executes the statistical processing of the history data.
This is an overlap command.
Syntax
:MEASure:HISTory:EXECute
Example
:MEASURE:HISTORY:EXECUTE
:MEASure:MODE
Function Sets the mode of the automated measurement of
waveform parameters or queries the current setting.
Syntax
:MEASure:MODE {BASic|CONTinuous|CYCLe|
HISTory}
:MEASure:MODE?
Example
:MEASURE:MODE BASIC
:MEASURE:MODE? -> :MEASURE:MODE BASIC
5.19 MEASure Group