5-55
IM DLM6054-17EN
Commands
1
2
3
4
5
6
App
Index
:ANALysis:WPARameter<x>:TRENd:
VERTical
Function Sets the vertical range of the trend of the waveform
parameter measurement or queries the current
setting.
Syntax
:ANALysis:WPARameter<x>:TRENd:VERTical
{<NRf>,<NRf>}
:ANALysis:WPARameter<x>:TRENd:VERTical?
<x> = 1 or 2
<NRf> = –1.000E+31 to 1.000E+31 (div)
Example
:ANALYSIS:WPARAMETER1:TREND:
VERTICAL 1,2
:ANALYSIS:WPARAMETER1:TREND:VERTICAL?
-> :ANALYSIS:WPARAMETER1:TREND:
VERTICAL 2.000E+00,1.000E+00
:ANALysis:XY<x>?
Function Queries all settings related to the XY display function.
Syntax
:ANALysis:XY<x>?
<x> = 1 or 2
:ANALysis:XY<x>:GATE?
Function Queries all settings related to the gate function of the
XY display.
Syntax
:ANALysis:XY<x>:GATE?
<x> = 1 or 2
:ANALysis:XY<x>:GATE:ALEVel
Function Sets the active level of the gate of the XY display or
queries the current setting.
Syntax
:ANALysis:XY<x>:GATE:ALEVel {HIGH|LOW}
:ANALysis:XY<x>:GATE:ALEVel?
<x> = 1 or 2
Example
:ANALYSIS:XY1:GATE:ALEVEL HIGH
:ANALYSIS:XY1:GATE:ALEVEL?
-> :ANALYSIS:XY1:GATE:ALEVEL HIGH
:ANALysis:XY<x>:GATE:HYSTeresis<x>
Function Sets the hysteresis of the gate of the XY display or
queries the current setting.
Syntax
:ANALysis:XY<x>:GATE:HYSTeresis<x>
{<NRf>}
:ANALysis:XY<x>:GATE:HYSTeresis<x>?
<x> of XY<x> = 1 or 2
<x> of HYSTeresis<x> = 1 to 8
<NRf> = 0 to 4 (div)
Example
:ANALYSIS:XY1:GATE:HYSTERESIS1 1
:ANALYSIS:XY1:GATE:HYSTERESIS1?
-> :ANALYSIS:XY1:GATE:
HYSTERESIS1 1.000E+00
:ANALysis:XY<x>:GATE:LEVel<x>
Function Sets the level of the gate of the XY display or queries
the current setting.
Syntax
:ANALysis:XY<x>:GATE:LEVel<x> {<NRf>,<V
oltage>,<Current>}
:ANALysis:XY<x>:GATE:LEVel<x>?
<x> of XY<x> = 1 or 2
<x> of LEVel<x> = 1 to 8
<NRf>, <Voltage>, and <Current> = See the DL6000/
DLM6000 User’s Manual.
Example
:ANALYSIS:XY1:GATE:LEVEL1 1
:ANALYSIS:XY1:GATE:LEVEL1?
-> :ANALYSIS:XY1:GATE:LEVEL1 1.000E+00
:ANALysis:XY<x>:GATE:TRACe
Function Sets the gate trace of the XY display or queries the
current setting.
Syntax
:ANALysis:XY<x>:GATE:TRACe {<NRf>|OFF}
:ANALysis:XY<x>:GATE:TRACe?
<x> = 1 or 2
<NRf> = 1 to 8
Example
:ANALYSIS:XY1:GATE:TRACE 1
:ANALYSIS:XY1:GATE:TRACE? -> :ANALYSIS:
XY1:GATE:TRACE 1
:ANALysis:XY<x>:MEASure?
Function Queries all settings related to the automated
measurement of the XY display.
Syntax
:ANALysis:XY<x>:MEASure?
<x> = 1 or 2
:ANALysis:XY<x>:MEASure:CURSor?
Function Queries all settings related to the cursor
measurement of the XY display.
Syntax
:ANALysis:XY<x>:MEASure:CURSor?
<x> = 1 or 2
:ANALysis:XY<x>:MEASure:CURSor:
XLINkage
Function Turns ON/OFF the ΔV cursor link on the XY display
or queries the current setting.
Syntax
:ANALysis:XY<x>:MEASure:CURSor:
XLINkage {<Boolean>}
:ANALysis:XY<x>:MEASure:CURSor:
XLINkage?
<x> = 1 or 2
Example
:ANALYSIS:XY1:MEASURE:CURSOR:
XLINkAGE ON
:ANALYSIS:XY1:MEASURE:CURSOR:XLINkAGE?
-> :ANALYSIS:XY1:MEASURE:CURSOR:
XLINkAGE 1
5.3 ANALysis Group