5-51
IM 701730-17E
Commands
5
5.11 GONogo Group
:GONogo:ZONE:ZONE<x>:SEND
Function
Sets the zone data specified for the GO/NO-GO
determination zone or queries the current
setting.
Syntax
:GONogo:ZONE:ZONE<x>:SEND
#6004028(data string)
:GONnogo:ZONE:ZONE<x>:SEND?
<x>=1 to 6 (1 to 3 on the DL1720E)
Example
:GONOGO:ZONE:ZONE1:SEND
#6004028(data string)
:GONOGO:ZONE:ZONE1:SEND? ->
:GONOGO:ZONE:ZONE1:SEND
#6004028(data string)
Description
• The (data string) consists of zone data of
length 4028 bytes.
• Zone data consist of 8-bit data (0 to 255).
This data string may contain ASCII codes
“0AH” or “00H” that stand for “NL.” Hence,
care must be taken when programming the
controller.
• This command is used to save the zone data
to the PC and set the data as zone data on
the instrument.
• When this command is used to set the zone
data to the instrument to perform the
determination, issue the
“
:GONogo:ZONE:ZONE<x>:MODE ON
”
command afterwards.
:GONogo:ZONE:ZONE<x>:TRACe
Function
Sets the target trace of the zone for GO/NO-GO
determination using zones or queries the
current setting.
Syntax
:GONogo:ZONE:ZONE<x>:TRACe
{<NRf>|MATH<y>}
:GONogo:ZONE:ZONE<x>:TRACe?
<NRf>=1 to 4 (1 to 2 on the DL1720E)
<x>=1 to 6 (1 to 3 on the DL1720E)
<y>=1 or 2 (1 on the DL1720E)
Example
:GONOGO:ZONE:ZONE1:TRACE 1
:GONOGO:ZONE:ZONE1:TRACE ->
:GONOGO:ZONE:ZONE1:TRACE 1