5-21
IM 701730-17E
Commands
5
:CHANnel<x>:MODE
Function
Turns ON/OFF the display of the channel or
queries the crrent setting.
Syntax
:CHANnel<x>:MODE {<Boolean>}
:CHANnel<x>:MODE?
<x>=1 to 4 (1 to 2 on the DL1720E)
Example
:CHANNEL1:MODE ON
:CHANNEL1:MODE? -> :CHANNEL1:MODE 1
Description
The “
CHANnel<x>:DISPlay
” command can
be used to make the same settings and
inquiries.
:CHANnel<x>:OFFSet
Function
Sets the offset voltage of the channel or queries
the current setting.
Syntax
:CHANnel<x>:OFFSet {<Voltage>}
:CHANnel<x>:OFFSet?
<x>=1 to 4 (1 to 2 on the DL1720E)
<Voltage>=See the User’s Manual
Example
:CHANNEL1:OFFSET 0V
:CHANNEL1:OFFSET? ->
:CHANNEL1:OFFSET 0.000E+00
Description
When a current probe is used, this command
sets or queries the <Current> value.
:CHANnel<x>:POSition
Function
Sets the vertical position of the channel or
queries the current setting.
Syntax
:CHANnel<x>:POSition {<NRf>}
:CHANnel<x>:POSition?
<x>=1 to 4 (1 to 2 on the DL1720E)
<NRf>=–4 to 4div (1/100 steps)
Example
:CHANNEL1:POSITION 1
:CHANNEL1:POSITION? ->
:CHANNEL1:POSITION 1.00E+00
:CHANnel<x>:PROBe
Function
Sets the probe attenuation/current-to-voltage
conversion ratio of the channel or queries the
current setting.
Syntax
:CHANnel<x>:PROBe {<NRf>|C10|C100}
:CHANnel<x>:PROBe?
<x>=1 to 4 (1 to 2 on the DL1720E)
<NRf>=1, 10, 100, or 1000 (see the User’s
Manual)
Example
:CHANNEL1:PROBE
:CHANNEL1:PROBE? ->
:CHANNEL1:PROBE 1
Description
When using the current probe, specify “C10”
(10 A:1 V) or “C100” (100 A:1 V), respectively.
:CHANnel<x>:VARiable
Function
Sets the voltage sensitivity of the channel in
0.01 steps of the V/div setting or queries the
current setting.
Syntax
:CHANnel<x>:VARiable {<Voltage>}
:CHANnel<x>:VARiable?
<x>=1 to 4 (1 to 2 on the DL1720E)
<Voltage>=See the Use’s Manual
Example
:CHANNEL1:VARIABLE 1V
:CHANNEL1:VARIABLE? ->
:CHANNEL1:VARIABLE 1.000E+00
Description
When a current probe is used, this command
sets or queries the <Current> value.
:CHANnel<x>:VDIV
Function
Sets the voltage sensitivity (V/div) of the
channel or queries the current setting.
Syntax
:CHANnel<x>:VDIV {<Voltage>}
:CHANnel<x>:VDIV?
<x>=1 to 4 (1 to 2 on the DL1720E)
<Voltage>=2 mV to 10V (when the probe
attenuation is 1:1, see the User’s Manual)
Example
:CHANNEL1:VDIV 2V
:CHANNEL1:VDIV? ->
:CHANNEL1:VDIV 2.000E+00
Description
When a current probe is used, this command
sets or queries the <Current> value.
5.5 CHANnel Group