5-49
IM 701730-17E
Commands
5
:GONogo:PARameter:ITEM<x>:TYPE:
<Parameter>
Function
Sets the upper and lower limits of the
automated measurement parameter of the
specified waveform parameter on which GO/
NO-GO determination is to be performed or
queries the current setting.
Syntax
:GONogo:PARameter:ITEM<x>:TYPE:
<Parameter>
{<{Voltage|DONTcare}>,<{Voltage|
DONTcare}>|<{Current|DONTcare}>,
<{Current|DONTcare}>|<{Time|
DONTcare}>,<{Time|DONTcare}>|
<{Frequency|DONTcare}>,<Frequency|
DONTcare}>|<{{<NRf>}|DONTcare}>,
<{{<NRf>}|DONTcare}>}
:GONogo:PARameter:ITEM<x>:TYPE:
<Parameter>?
<Parameter>=
{AVERage|AVGFreq|
AVGPeriod|BWIDth1|BWIDth2|DELay|
DUTYcycle|FALL|FREQuency|HIGH|LOW|
MAXimum|MINimum|NOVershoot|NWIDth|
PERiod|PNUMber|POVershoot|PTOPeak|
PWIDth|RISE|RMS|SDEViation|
TY1Integ|TY2Integ|XY1Integ|XY2Integ}
<x>=1 to 4
<Voltage>, <Time>, <Frequency>, <NRf>=See
the User’s Manual.
Example
:GONOGO:PARAMETER:ITEM1:TYPE:
AVERAGE 100mV,-100mV
:GONOGO:PARAMETER:ITEM1:TYPE:
AVERAGE? ->
:GONOGO:PARAMETER:ITEM1:
TYPE:AVERAGE 100.0E-03,-100.0E-03
Description
• Queries are only allowed on the parameters
that have been set using this command.
• When a current probe is used, this command
sets or queries the <Current> value.
:GONogo:PARameter:ITEM<x>:VALue?
Function
Queries the automated measurement values of
the waveform parameters on which GO/NO-GO
determination is to be performed.
Syntax
:GONogo:PARameter:ITEM<x>:VALue?
<x>=1 to 4
Example
:GONOGO:PARAMETER:ITEM1:VALUE? ->
:GONOGO:PARAMETER:ITEM1:
VALUE 1.98E-03
Description
If the measurement is not possible (the mode is
OFF, for example), “NAN(Not A Number)” is
returned.
:GONogo:SEQuence
Function
Selects single or continuous action or queries
the current setting.
Syntax
:GONogo:SEQuence
{CONTinuous|SINGle}
:GONogo:SEQuence?
Example
:GONOGO:SEQUENCE SINGLE
:GONOGO:SEQUENCE? ->
:GONOGO:SEQUENCE SINGLE
:GONogo:STOPcount
Function
Sets the number of waveform acquisitions on
which to perform GO/NO-GO determination or
queries the current setting.
Syntax
:GONogo:STOPcount {<NRf>|INFinity}
:GONogo:STOPcount?
<NRf>=1 to 65536
Example
:GONOGO:PARAMETER:STOPCOUNT 10
:GONOGO:PARAMETER:STOPCOUNT? ->
:GONOGO:STOPCOUNT 10
:GONogo:TLINkage
Function
Turns ON/OFF the linked movement of the
determination range cursor of the GO/NO-GO
determination.
Syntax
:GONogo:TLINkage {<Boolean>}
:GONogo:TLINkage?
Example
:GONOGO:TLINKAGE ON
:GONOGO:TLINKAGE? ->
:GONOGO:TLINKAGE 1
:GONogo:TRANge
Function
Sets the range over which to perform GO/NO-
GO determination or queries the current setting.
Syntax
:GONogo:TRANge {<NRf>},{<NRf>}
:GONogo:TRANge?
<NRf>=–5.00 to 5.00 (10 div/displayed record
length steps)
Example
:GONOGO:TRANGE -5.00,5.00
:GONOGO:TRANGE? ->
:GONOGO:TRANGE -5.00,5.00
:GONogo:WAIT?
Function
Waits for the GO/NO-GO determination to
complete with a timeout.
Syntax
:GONogo:WAIT? {<NRf>}
<NRf>=1 to 864000 (in units of 100 ms)
Example
(Set the timeout to 5 s.)
:GONOGO:WAIT? 50 -> :GONOGO:WAIT 0
Description
Returns “1” if it times out, “02 if it completes
before timing out.
:GONogo:ZONE?
Function
Queries all settings related to GO/NO-GO
determination using zones.
Syntax
:GONogo:ZONE?
Example
:GONOGO:ZONE? -> :GONOGO:ZONE:
ZONE1:TRACE 1;MODE 0;
CONDITION OUT;:GONOGO:ZONE:ZONE2:
TRACE 2;MODE 0;CONDITION OUT;:
GONOGO:ZONE:ZONE3:TRACE 3;MODE 0;
CONDITION OUT;:GONOGO:ZONE:ZONE4:
TRACE 4;MODE 0;CONDITION OUT;:
GONOGO:ZONE:ZONE5:TRACE MATH1;
MODE 0;CONDITION OUT;:GONOGO:ZONE:
ZONE6:TRACE MATH2;MODE 0;
CONDITION OUT
5.11 GONogo Group