IM CA71-E
7-5
Functions Provided by DIP
Switch
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Step 8:
Using each pair of
and
keys, set the value of the
denominator
m
and the starting setpoint of the enumerator
n
.
(See Section 4.6, "Divided Output Function (
n/m
), for further
details.) The starting setpoint is the minimum of the variable
n
for auto step operation.
Step 9:
Pressing the
key initiates the automatic stepping of the
divided output (
n/m
), as described below.
Assuming the starting value of the variable
n
is 1, the output
cyclically changes with the variable
n
as
n
changes in the
following manner.
n
= 1
→
2
→
3
→
· · · (
m
– 1)
→
m
→
(
m
– 1)
→
· · ·
→
2
→
1
→
2
→
3
→
· · ·
Step 10:
To quit the auto step function, turn off the calibrator by
pressing the
key. Place switch 1 back in the OFF
(left-side) position. Note that disabling the divided output
mode (
n/m
) before placing switch 1 back in the OFF position
enters sweep mode.
TIP
• Pressing the
key to change to the SOURCE OFF state pauses the
auto step operation.
• To execute the auto step function again, press the
key.
The auto step operation resumes from the value of n shown on the LCD.
7.2 Auto Step Function