5.1 Connecting Cables to Terminals
CAUTION
• Before connecting the calibrator to the device under test, cut off the power
to the device.
• Do not apply any voltage or current exceeding the allowable voltage (300 V)
or current (120 mA). Otherwise, there will be a danger of not only damage
to the instrument but also personal injury due to electrical shock.
• Mistaking the H voltage input terminal for the mA current input terminal,
and vice versa, when wiring is extremely dangerous.
NEVER make this mistake.
• The current input terminals are equipped with a built-in current input
protection fuse. Overcurrent input to the terminals will cause the fuse to blow.
If the fuse is blown, replace it with one (P/N: A1635EF) with the specified
ratings.
For details on fuse replacement, see subsection 5.2.3, "Measuring DC
Current."
• When using the terminal adapter (model: 99021), tighten the knob by hand.
Do not use a tool or the like. Tightening the knob using a tool or the like
may damage the terminal, resulting in the disability of measurement.
Before storing the instrument in the carrying case, remove the terminal
adapter (model: 99021). If the instrument is stored in the carrying case
while the adapter is attached, an external force may be applied to the
terminal, thus causing damage to the terminal and resulting in the
disability of measurement.