IM CA71-E
8-1
Communication Function
1
2
3
4
5
6
7
8
9
10
11
12
13
App
8.
Communication Function - CA71 only -
You can configure the calibrator from a personal computer just as
you do with the calibrator's panel keys (except for turning on/off
the power, configuring the function selector switch, and setting the
communication function). You can also verify the setpoint, measured
value and status of the calibrator.
NOTE
• With the optional communication cable (P/N: 91017), you can use the RS232
serial port (9-pin D-sub) of a personal computer or any other equipment.
• In talk-only or printer mode, you can output the source setpoint and
measured value at preset intervals.
8.1 Cables Connection and Interface
Specifications
■ Connecting Communication Cable
Remove the I/O port cover at the back of the calibrator and connect
the communication cable (P/N: 91017) to the I/O port.
■ Setting RS232 Parameters
Baud rate:
9600 baud
Parity: None
Stop bits:
2 bits
Data length:
8 bits
Flow control: None (Xon/Xoff control for printing only)
Delimiter:
Fixed to CrLf