4-8
IM CA71-E
■ Output Method Based on Three-wire Connection
Attach another lead cable to the L output terminal, as shown in the
following figure. The output is provided through the three wires,
H, L and L’. Connect these three wires to the device being calibrated.
CA71
SOURCE
Three-wire measuring equipment
H
H
L
L'
L
Three-wire Connection for Resistance Signal Source
Step 1:
Using the function selector switch, select
.
Step 2:
Using the
key, select the range. Pressing the
key
cycles through the 400 Ω, PT100 and JPT100 options.
Step 3:
Set the output value digit by digit using each pair of
and
keys. Each press of the
or
key increases or
decreases the digit. Increasing the digit from 9 or decreasing
it from 0 causes the digit to overflow or underflow, allowing
you to set the output value without interruption. Holding down
the
or
key continuously changes the digit in question.
Pressing the
key initializes the output setpoint to the
default value (0).
Step 4:
Pressing the
key causes the
indicator on the LCD
to change from
to
. The calibrator sources the preset
resistance value between the output terminals.
Step 5:
To turn off the output, press the
key once again.
The
appears on the LCD and the output terminals are
open-circuited.
4.3 Sourcing Resistance or RTD Signal