5-10
IM AQ6380-02EN
No. Message
Cause of Warning
108 Marker setting out of range
An attempt was made to execute the analysis function between
markers in the state where both line marker 1 and line marker 2
were set outside the measurement range.
109 Auto sweep failed
Although the AUTO sweep started, the sweep stopped because
optimum conditions were not found.
110 No data between line markers
An attempt was made to execute the analysis function in a state
where there was no data in the line markers of an active trace.
111 <G=MKR FIT> failed
Trace G cannot write fitting curve because numbers of data is not
sufficient during <G=MKR FIT>.
120 USB storage not found
USB storage medium not inserted.
121 USB storage not initialized
SUB storage not initialized.
122 USB storage write protected
The USB storage device is write protected.
123 File not found
The specified file cannot be read because it has not been found. Or,
the file does not exist on the disk.
124 Improper directory name
Directory cannot be created because the name is not valid.
125 Improper file name
A file cannot be saved due to an incorrect file name.
126 Directory already exists
Directory cannot be created because a directory of the same name
already exists.
128 A write-protected file
The file cannot be re-written or deleted because its property has
been set to READ ONLY.
129 USB Storage full
The file cannot be saved, because the hard disk or the USB storage
device is full.
130 Directory full
No files can be created because the directory area is full.
131 No data
There was no data to be saved, although an attempt was made to
save the file.
132 Trace file not found
The file cannot be read because it is not a waveform file.
133 Program file not found
The file cannot be read because it is not a program file.
134 Data file not found
The file cannot be read because it is not a data file.
135 Setting file not found
The file cannot be read because it is not a settings file.
138 Copy denied for invalid file name The file copy cannot be executed because the “copy from” file and
the “copy to” file have the same name.
140 Paste denied for lack of line space The paste operation cannot be executed during the editing of
programs because the number of blank rows is not sufficient.
141 Denied for limit of lines maximum It is not possible to execute a merge during the editing of programs
because the merged result would exceed the maximum number of
rows.
142 WL calibration failed
Calibration cannot be executed because the level of a light source
is not sufficient during the wavelength calibration or because
wavelength deviations exceed the calibration range.
143 Optical alignment failed
Alignment adjustment cannot be executed because the level of a
light source is not sufficient during the alignment adjustment.
147 RBW calibration failed (Power)
Rms resolution calibration cannot be executed because the light
source level is too low.
148 RBW calibration failed (Spectrum) Rms resolution calibration cannot be executed because the light
source spectral width is too wide.
149 RBW calibration failed (Stability)
Rms resolution calibration cannot be executed because the light
source level is unstable.
154 Incompatible file type
The file could not be loaded because the model does not support it.
170 Invalid character
An illegal character was entered for a network name.
171 Invalid address
An illegal address was set to an IP address.
172 Incompatible firmware version
Attempted to load an incompatible update file.
173 Update file read error
An incorrect number of files or file corruption was found when
attempting to load the update file.
174 Invalid password
The password entered for the key lock is not correct.
175 Re-enterd password unmatched
The two passwords entered for the key lock do not match.
176 Enter 4-digit number for password The password entered for the key lock is not a 4-digit number.
177 Invalid port number
An attempt was made to set the port number for remote control to
an invalid number (1025, 20001, 30001).
5.7 Warning Display Function