3-14
IM AQ6380-02EN
3.4 Alignment Adjustment
Procedure
1.
Turn the power to the instrument ON. Turn ON the
MAIN POWER
switch and press the
POWER
switch. For instructions on turning the power ON and OFF, see section 3.3.
2.
Tap System. The system menu is displayed.
3.
Tap the
Optical Alignment
.
4.
Tap the
Execute
. Alignment adjustment is executed automatically by using the alignment light
source built into the instrument. Adjustment finishes a few minutes thereafter, and you are
returned to the original screen.
6.
Tap
Cancel
to cancel alignment adjustment partway through the process.
Note
• After alignment has been executed, wavelength calibration is also performed automatically inside the
instrument.
• The alignment adjustment has no effect if the process is stopped. The instrument remains in the state
prior to execution of the alignment adjustment.