58
Reference Value Out of Range Error
Displayed when the measured value used for
setting or updating the relative measurement
reference value is outside the valid range.
Exceeds the upper limit: High level error.
Falls below the lower limit: Low level error.
Low Battery Error
Displayed when attempting to save or delete
data while the battery indicator is displayed.
The error appears for one second, then the
screen returns to the previous screen.
CAL Value Automatic Setting Error
1) Displayed when attempting to automatically
set the CAL value a second time during
relative value measurement. Clear the
CAL value (see section 3-6-3).
2) Displayed when the instrument attempts to
automatically set a CAL value outside the
valid range (-10.00 dB–25.00 dB).
Change the fixed optical attenuator to one
having an amount of attenuation that will
bring the CAL value into the valid range.
Resume Error
If the Resume value was not saved
successfully, the Resume error appears the
next time the instrument is started. Press the
ENTER key for normal measurement mode.
Or, turn OFF the power switch and restart.
The system reset initial value is set (see
section 2-5-5).
Calibration Value Error
Displayed when, for some reason, the
instrument's internal calibration value
becomes corrupted during use.
Have the instrument repaired immediately.
nm
CW
PWR SAVE
nm
CW
PWR SAVE
PWR SAVE
kHz
nm
PWR SAVE
kHz
nm
PWR SAVE
PWR SAVE
PWR SAVE
B
PWR SAVE
PWR SAVE
B
B
PWR SAVE
PWR SAVE
PWR SAVE
Содержание AQ2160-02
Страница 1: ...User s Manual IM SU1005A 01E 3rd Edition Model SU1005A AQ2160 02 Optical Powermeter ...
Страница 71: ...63 Outline drawing of the AQ2160 02 Optical Powermeter with the SU2004A SCC connector adapter installed ...
Страница 72: ...64 SCC LCC FCC MUC STC Outline drawing of the SU2004A connector adapter ...