8
1-7. Specifications
Specifications of the AQ2160-02 Optical Powermeter
Wavelength range
750 nm–1700 nm (can be set in 5 nm steps)
Sensor element
InGaAs (1 mm
φ
)
Compatible fibers
SM (9/125 µm), GI (50/125 µm), GI (62.5/125 µm)
fibers
Input type
1
Connector adapter (optional)
Power range
2
-70 dBm–+10 dBm
Noise level
2
-60
dBm
Accuracy
3
±5%
Measured value
display
7-segment, 4-digit display with backlight
Displayed units
Absolute values: dBm, mW, µW, nW
Relative values: dB
Range switching
Automatic
Measurement
mode
Select CW light, or chopped light (270 Hz, 1 kHz, 2
kHz)
Measurement
interval
Approximately 330 ms
Display resolution
0.01 dB
Relative value
measurement
Relative value measurement relative to a set
reference value. Relative value measurement
based on a displayed measurement.
Backlight
Illuminates while the Backlight key is pressed, and
stays on for approximately five seconds thereafter.
Power saving
Power save function turns the power OFF
automatically if no key is pressed for ten minutes
(function can be turned OFF).
Battery check
Low battery indicator
Resume function
4
Restores the settings active when the power was
last turned OFF.
Memory function
Measured values, reference values, or
measurement conditions can be saved or deleted to
or from memory: 1000 (standard memory) + 100
(user memory).
Interface
5
USB version 1.1
Содержание AQ2160-02