46
6.
Maintenance Method
Replacing the deaeration membrane module
1. Turn OFF the earth leakage breaker of the unit.
2. Close the tap.
3. Open the door of the main body. (See Photo 1)
Photo
1
Photo 2
4. Remove the pre-treatment cartridge and ion exchange resin (CPC-N).
5. Remove a plate fixing the ion exchange resin (CPC-N).
(
See Photos 2 and 3
)
Photo 3
6. Remove the piping connected to the deaeration membrane module as well as fixing
screws.
7. For installation, reverse the above procedure.
TOC reduction UV lamp unit
Deaeration menbrance module
Vaccum pump for deaeration
menbrance module