Relay outputs
System 182-XT-4
6 - 24
ba76027e02
12/2014
Canceling the cleaning
A running cleaning cycle is canceled:
Automatically
– if the sensor switches to the inactive condition during the cleaning
cycle
Manually
– By pressing
c
– By switching on the maintenance condition
Each time the cleaning cycle is canceled, the relay opens immediately.
If the cleaning cycle is canceled automatically, the outputs linked to the
sensor are released immediately.
If the cleaning cycle is canceled manually, the sensor is in the
maintenance condition. The linked outputs are only released after the
maintenance condition was terminated manually.
The next cleaning cycle will be performed at the time set up.
6.4.9
Manual control
Function
The Manual control function can be used to test the operativeness of
an instrument that is connected to the relay. To do so, close or open the
relay manually and, while doing so, check the behavior of the
connected instrument.
In case of a power failure, all relays open. The cleaning
cycle is canceled. The outputs linked with the sensor
change to the non-active condition (see section 6.5.2). As
soon as the power is available, the outputs are released
again. The next cleaning cycle will be performed at the time
set up.
Setting
Selection/Values
Explanation
Relay function
Manual control
The selected relay action
is carried out with Save
and quit.
Action
Open
Close
Relay action
The settings for other functions in the Relay function menu
as, for example, Frequency output and Pulse-width output
are retained while the Manual control is carried out.
Содержание YSI 182-XT-4
Страница 6: ...Contents System 182 XT 4 0 4 ba76027e02 12 2014...
Страница 10: ...Overview System 182 XT 4 1 4 ba76027e02 12 2014...
Страница 14: ...Safety System 182 XT 4 2 4 ba76027e02 12 2014...
Страница 64: ...Installation System 182 XT 4 3 50 ba76027e02 12 2014...
Страница 82: ...Operation System 182 XT 4 4 18 ba76027e02 12 2014...
Страница 88: ...The SETTINGS menu System 182 XT 4 5 6 ba76027e02 12 2014...
Страница 114: ...Relay outputs System 182 XT 4 6 26 ba76027e02 12 2014...
Страница 116: ...Maintenance and cleaning System 182 XT 4 7 2 ba76027e02 12 2014...
Страница 122: ...What to do if System 182 XT 4 8 6 ba76027e02 12 2014...
Страница 134: ...Technical data System 182 XT 4 9 12 ba76027e02 12 2014...
Страница 135: ...System 182 XT 4 1 ba76027e02 12 2014...
Страница 138: ...Accessories and options System 182 XT 4 11 2 ba76027e02 12 2014...
Страница 142: ...Appendix store separately if required System 182 XT 4 13 2 ba76027e02 12 2014...
Страница 143: ......