System 182-XT-4
Relay outputs
6 - 23
ba76027e02
12/2014
Example
Fig. 6-12 Cleaning cycle
Setting
Result
Reference time (h):
Reference time
(min):
Interval unit:
Cleaning interval:
12
0
Hours (h)
8 h
Reference time: 12:00 hours
This specifies the following
start times:
04:00, 12:00 and 20:00 hours
1
Reference time
Start of a cleaning cycle (t1)
Start of the specified Cleaning duration (t1a)
2
End of the specified Cleaning duration (t1a)
Start of the specified Adjustment time (t1b)
3
End of the specified Adjustment time (t1b)
End of the cleaning cycle (t1)
4
Reference time ± Cleaning interval (t2)
Start of a cleaning cycle
t1
Cleaning cycle =
Cleaning duration (t1a) Adjustment time (t1b)
Linked outputs are frozen.
t2
Cleaning interval
relay
condition
open
closed
00:00
Time
3
1
2
t1
t1
t1
t1a t1b
4
4
t2
t2
t2
t2
1
04:00
12:00
24:00
20:00
Содержание YSI 182-XT-4
Страница 6: ...Contents System 182 XT 4 0 4 ba76027e02 12 2014...
Страница 10: ...Overview System 182 XT 4 1 4 ba76027e02 12 2014...
Страница 14: ...Safety System 182 XT 4 2 4 ba76027e02 12 2014...
Страница 64: ...Installation System 182 XT 4 3 50 ba76027e02 12 2014...
Страница 82: ...Operation System 182 XT 4 4 18 ba76027e02 12 2014...
Страница 88: ...The SETTINGS menu System 182 XT 4 5 6 ba76027e02 12 2014...
Страница 114: ...Relay outputs System 182 XT 4 6 26 ba76027e02 12 2014...
Страница 116: ...Maintenance and cleaning System 182 XT 4 7 2 ba76027e02 12 2014...
Страница 122: ...What to do if System 182 XT 4 8 6 ba76027e02 12 2014...
Страница 134: ...Technical data System 182 XT 4 9 12 ba76027e02 12 2014...
Страница 135: ...System 182 XT 4 1 ba76027e02 12 2014...
Страница 138: ...Accessories and options System 182 XT 4 11 2 ba76027e02 12 2014...
Страница 142: ...Appendix store separately if required System 182 XT 4 13 2 ba76027e02 12 2014...
Страница 143: ......