System 182-XT-4
Installation
3 - 33
ba76027e02
12/2014
– marked as an interrupt facility for the DIQ/S 182-XT-4-PR.
After it has been installed, the DIQ/S 182-XT-4-PR may only be
opened if the mains voltage has been switched off beforehand.
Materials required
Wire end sleeves, suitable for the power line, with suitable crimping
tool
1 x cable gland, matched to the cable diameter (see section 3.7 on
page 3-31).
Tools
Cable stripping knife
Wire stripper
Phillips screwdriver
Small screwdriver.
Preparing the power
cable
Fig. 3-25 Prepared power cable.
NOTE
The ground wire must not project into the enclosure. Otherwise, mal
-
functions could occur.
1
Cut off the cable to the required length.
2
Strip the cable insulation for approx. 45 mm.
3
Bare the wires of phases L and N and fit them with wire end
sleeves.
4
If present, cut off the ground wire at the end of the cable
sheath.
Содержание YSI 182-XT-4
Страница 6: ...Contents System 182 XT 4 0 4 ba76027e02 12 2014...
Страница 10: ...Overview System 182 XT 4 1 4 ba76027e02 12 2014...
Страница 14: ...Safety System 182 XT 4 2 4 ba76027e02 12 2014...
Страница 64: ...Installation System 182 XT 4 3 50 ba76027e02 12 2014...
Страница 82: ...Operation System 182 XT 4 4 18 ba76027e02 12 2014...
Страница 88: ...The SETTINGS menu System 182 XT 4 5 6 ba76027e02 12 2014...
Страница 114: ...Relay outputs System 182 XT 4 6 26 ba76027e02 12 2014...
Страница 116: ...Maintenance and cleaning System 182 XT 4 7 2 ba76027e02 12 2014...
Страница 122: ...What to do if System 182 XT 4 8 6 ba76027e02 12 2014...
Страница 134: ...Technical data System 182 XT 4 9 12 ba76027e02 12 2014...
Страница 135: ...System 182 XT 4 1 ba76027e02 12 2014...
Страница 138: ...Accessories and options System 182 XT 4 11 2 ba76027e02 12 2014...
Страница 142: ...Appendix store separately if required System 182 XT 4 13 2 ba76027e02 12 2014...
Страница 143: ......