Operation
pHotoFlex
®
Turb
74
ba75976e07
09/2021
A software update comprises
new instrument software
new programs (methods)
revisions of existing methods
The current software version can be found on the Internet at
www.WTW.com
.
The proceeding for updating the software is given in the appendix (see
section 11).
4.13 Administrating user-defined methods
User-defined programs can be
entered
read out
erased.
To store user-defined programs (methods) in the meter, determine a calibra-
tion line for your method yourself and transmit the data of this calibration line
to the pHotoFlex
®
Turb.
Administration of the method data can be carried out in the LSdata software
(see LSdata operating manual) or with the aid of a terminal program ( see
section 4.13.1).
You need the following data in any case:
Data
Possible entries/examples
Program No
:
900 ... 999
Model
:
Any name (max. 5 characters)
Cuvette [mm]
:
16 or 28
Wave length [nm]
:
436, 517, 557, 594, 610 or 690
MRB
:
(measuring range beginning)
e.g. 0.05
MRE
:
(measuring range end)
e.g. 8.00
E0
: (Offset)
e.g. 0.0
Slope
: (slope)
e.g. 1.0
Resolution
: (resolution)
0.0001, 0.001, 0.01, 0.1, 1, 2, 5, 10 or 100
Formula
: (citation form)
e.g. PO4-P (max. 9 characters)