pHotoFlex
®
Turb
Operation
71
ba75976e07
09/2021
Example,
output format CSV
The CSV output format delivers datasets separated by ";".
The data are output in the following order:
1 Current number of the storage location
(or "0" during output from measured value display)
2 Date of storing
3 Time of storing
4 Selected ID
5 Program number (
only for the measured parameter,
Concentra-
tion
)
6 Citation form (
only for the measured parameter,
Concentration
)
7 AR:
Measuring mode
pH & ORP
: AutoRead
BV:
Measured parameter
Concentration
: Blank value
8 Dilution (only for the measured parameter,
Concentration
)
9 Main measured value or
Upper/lower measuring range limit
(only with measured value status, OFL/UFL)
10Unit of the main measured value
or
designation of dimensionless measured values, e.g. <pH>
11Measured value status of the main measured value
* VALID: Measured value valid
* INVALID: Measured value invalid
* UFL: Measured value below the lower measuring range limit
* OFL: Measured value above the upper measuring range limit
12
Secondary measured value
with measured parameter,
pH & ORP
: Temperature
with measured parameter,
Absorbance
/
% Transmission
:
Wavelength of measurement
13Unit of the secondary measured value
for the measured parameter,
pH & ORP
: °C / °F
with measured parameter,
Absorbance
/
% Transmission
: nm
14Measured value status of the secondary measured value (see 11)
If the connected external printer does not print, please check
whether the same baud rate is set at the pHotoFlex
®
Turb and
printer.
0;01.12.11;14:18:07;1;81;O2;[BV];[2 + 1];166;mg/l;VALID;;;
14;03.12.11;07:41:51;3;;;;;24.8;%;VALID;557;nm;VALID
2
1
3
4 5
6
7
8
9
12
10
13
11
14