![Xylem Multi 3620 IDS Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/xylem/multi-3620-ids/multi-3620-ids_operating-manual_3922862069.webp)
Multi 3620 IDS
Turbidity measurement (VisoTurb
®
900-P)
ba77169d06
10/2021
69
Calibration evalua-
tion
After calibrating, the meter automatically evaluates the calibration.
Calibration record
(USB output)
Menu item
Setting/
function
Explanation
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<PRT>
.
Output all calibration records to
the interface with
<PRT_>
.
Quit the display with
<ESC>
or
<MENU/ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration records to
the interface.
Display
Calibration record
Explanation
+++
Optimum calibration
Good calibration
Multi 3620 IDS
Ser. no. 12345678
CALIBRATION TRB:
VisoTurb 900-P
Ser. no. 14E999003
18.09.2014 08:09:10
# 1
0.0 FNU
# 2
124.0 FNU
Sensor
+++
_____________________________________
Содержание Multi 3620 IDS
Страница 1: ...OPERATING MANUAL ba77169d06 10 2021 Multi 3620 IDS DIGITAL METER FOR IDS SENSORS 3620 IDS...
Страница 2: ...Multi 3620 IDS 2 ba77169d06 10 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3620 IDS Commissioning ba77169d06 10 2021 15 3 3 4 Setting the date and time 1 See section 4 5 5...
Страница 110: ...Glossary Multi 3620 IDS 110 ba77169d06 10 2021...
Страница 113: ......