Multi 3620 IDS
pH value
ba77169d06
10/2021
35
3.
Thoroughly rinse the IDS sensor with deionized water.
4.
Immerse the IDS-pH sensor in buffer solution 1.
5.
For measurements without temperature sensor
(e.g. when using an IDS adapter
)
:
Measure the temperature of the buffer manually and enter it with
<
><
>
.
6.
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
7.
Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<MENU/ENTER>
.
The pH value of the buffer solution is displayed.
15.03.2014 08:00
ConCal
15.03.2014 08:00
ConCal