What to do if...
Multi 3620 IDS
102
ba77169d06
10/2021
Obviously incorrect
measured values
15.3
Dissolved oxygen
Error message,
OFL
Error message,
Error
15.4
Conductivity
Error message,
OFL
Error message,
Error
Cause
Remedy
IDS-pH sensor:
– IDS-pH sensor unsuitable
– Use suitable IDS sensor
– Temperature difference
between buffer and test sample
too great
– Adjust temperature of buffer or sam-
ple solutions
– Measurement procedure not
suitable
– Follow special procedure
Cause
Remedy
– Measured value outside the
measuring range
– Select a suitable measuring medium
Cause
Remedy
– Sensor contaminated
– Clean the sensor
– Measured temperature value
outside the operating conditions
(display of OFL/UFL instead of a
temperature value)
– Keep to the temperature range for
the test sample
– Defective sensor
– Replace the sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– Measured value outside the
measuring range
– Use suitable IDS conductivity sensor
Cause
Remedy
– Sensor contaminated
– Clean the sensor and replace it if
necessary
– Calibration solution not suitable
– Check the calibration solutions
Содержание Multi 3620 IDS
Страница 1: ...OPERATING MANUAL ba77169d06 10 2021 Multi 3620 IDS DIGITAL METER FOR IDS SENSORS 3620 IDS...
Страница 2: ...Multi 3620 IDS 2 ba77169d06 10 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3620 IDS Commissioning ba77169d06 10 2021 15 3 3 4 Setting the date and time 1 See section 4 5 5...
Страница 110: ...Glossary Multi 3620 IDS 110 ba77169d06 10 2021...
Страница 113: ......