Turbidity measurement (VisoTurb
®
900-P)
Multi 3620 IDS
64
ba77169d06
10/2021
Selecting the
displayed
measured parameter
You can switch between the following displays with
<M>
:
Turbidity [FNU]
Turbidity [NTU]
Freezing the mea-
sured value (HOLD
function)
With the HOLD function, you can freeze the current measured value. The dis-
played measured value stops changing until you switch the HOLD function off.
Stability control
(AutoRead )
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values. The display of the measured parameter flashes until
a stable measured value is available.
You can start the
Stability control
function manually at any time, irrespective of
the setting for automatic
Stability control
(see section 11.6.3) in the
System
menu.
30.09.2012 08:00
Info
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
If the HOLD function is active, you can, e.g. start a manual mea-
surement with stability control.
2.
Release the frozen measured value again with
<AR>
.
The HOLD function is switched off.
The [HOLD] status display disappears.
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
Содержание Multi 3620 IDS
Страница 1: ...OPERATING MANUAL ba77169d06 10 2021 Multi 3620 IDS DIGITAL METER FOR IDS SENSORS 3620 IDS...
Страница 2: ...Multi 3620 IDS 2 ba77169d06 10 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3620 IDS Commissioning ba77169d06 10 2021 15 3 3 4 Setting the date and time 1 See section 4 5 5...
Страница 110: ...Glossary Multi 3620 IDS 110 ba77169d06 10 2021...
Страница 113: ......