Multi 3620 IDS
What to do if...
ba77169d06
10/2021
103
15.5
Turbidity
Implausible turbidity
values
Error message,
OFL
Measured values too
low
Measured values too
high
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– There are gas bubbles (e.g. air
bubbles) in front of the measure-
ment window
– Remove the gas bubbles, e.g.
by immersing the sensor at an angle
– Incorrect calibration,e.g.:
– Unsuitable calibration stan-
dard solutions (e.g. too old)
– Unsuitable calibration envi-
ronment (e.g. gas bubbles,
reflection, light)
– Check the calibration
– Minimum depth of immersion not
observed
– Heed the minimum depth of immer-
sion of the sensor (2 cm)
Cause
Remedy
– Measured value outside the
measuring range
– Select a suitable measuring medium
Cause
Remedy
– Measurement window soiled
– Clean the measurement window
Cause
Remedy
– Reflection at the walls or bottom
of the measuring vessel
– Keep the minimum distance of the
sensor towards the walls and bot-
tom of the measuring vessel (see
section 15.5)
– Incidence of light
– Use a light-proof measuring vessel
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Содержание Multi 3620 IDS
Страница 1: ...OPERATING MANUAL ba77169d06 10 2021 Multi 3620 IDS DIGITAL METER FOR IDS SENSORS 3620 IDS...
Страница 2: ...Multi 3620 IDS 2 ba77169d06 10 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3620 IDS Commissioning ba77169d06 10 2021 15 3 3 4 Setting the date and time 1 See section 4 5 5...
Страница 110: ...Glossary Multi 3620 IDS 110 ba77169d06 10 2021...
Страница 113: ......