Multi 3620 IDS
pH value
ba77169d06
10/2021
33
Continuing with two-
point calibration
Continuing with
three- to five-point
calibration
9.
Thoroughly rinse the IDS sensor with deionized water.
10.
Immerse the pH sensor in buffer solution 2.
11.
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it with
<
><
>
.
12.
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
13.
Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage display).
14.
If necessary, finish the calibration procedure as a two-point calibration
with
<M>
.
The calibration record is displayed.
30.09.2012 08:00
15.
Thoroughly rinse the IDS-pH sensor with deionized water.
16.
Immerse the IDS-pH sensor in the next buffer solution.
17.
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it with
<
><
>
.
18.
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
Содержание Multi 3620 IDS
Страница 1: ...OPERATING MANUAL ba77169d06 10 2021 Multi 3620 IDS DIGITAL METER FOR IDS SENSORS 3620 IDS...
Страница 2: ...Multi 3620 IDS 2 ba77169d06 10 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3620 IDS Commissioning ba77169d06 10 2021 15 3 3 4 Setting the date and time 1 See section 4 5 5...
Страница 110: ...Glossary Multi 3620 IDS 110 ba77169d06 10 2021...
Страница 113: ......