![Xylem Multi 3620 IDS Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/xylem/multi-3620-ids/multi-3620-ids_operating-manual_3922862067.webp)
Multi 3620 IDS
Turbidity measurement (VisoTurb
®
900-P)
ba77169d06
10/2021
67
Continuing with two-
point calibration
6.
Thoroughly rinse the turbidity sensor with distilled water and dry it with
a lint-free cloth.
7.
Immerse the turbidity sensor in the test sample at an angle.
8.
Position the turbidity sensor in the measuring vessel.
9.
Use
<
V
> <
W
>
and
<
><
>
to set the concentration of the standard
solution for each digit and confirm with
<MENU/ENTER>
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
10.
Wait for the end of the AutoRead measurement.
The calibration display for the next standard solution appears.
15.09.2014 08:00
0.0 FNU
Standard
1
Cal
15.09.2014 08:00
124.0 FNU
Standard
2
Cal
11.
Thoroughly rinse the turbidity sensor with distilled water and dry it with
a lint-free cloth.
12.
Immerse the turbidity sensor in the test sample at an angle.
13.
Position the turbidity sensor in the measuring vessel.
14.
Use
<
V
> <
W
>
and
<
><
>
to set the concentration of the standard
solution for each digit and confirm with
<MENU/ENTER>
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
15.
Wait for the end of the AutoRead measurement.
The calibration display for the next standard solution appears.
Содержание Multi 3620 IDS
Страница 1: ...OPERATING MANUAL ba77169d06 10 2021 Multi 3620 IDS DIGITAL METER FOR IDS SENSORS 3620 IDS...
Страница 2: ...Multi 3620 IDS 2 ba77169d06 10 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3620 IDS Commissioning ba77169d06 10 2021 15 3 3 4 Setting the date and time 1 See section 4 5 5...
Страница 110: ...Glossary Multi 3620 IDS 110 ba77169d06 10 2021...
Страница 113: ......