Multi 3620 IDS
Conductivity
ba77169d06
10/2021
61
Calibration evalua-
tion
After calibrating, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Menu item
Setting/func-
tion
Explanation
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<PRT>
.
Output all calibration records to
the interface with
<PRT_>
.
Quit the display with
<ESC>
or
<MENU/ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data
storage
/
Output to
USB flash drive or
printer
-
Outputs the stored calibration data
to the USB-A interface (
USB Host
)
(USB memory device/USB printer)
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the stored calibration data
to the USB-B interface (
USB
Device
) (PC)
Display
Calibration
record
Cell constant [cm
-1
]
+++
Within the range
0.450 ... 0.500 cm
-1
Error
Error
Outside the range
0.450 ... 0.500 cm
-1
Perform error elimination according to chapter 15 W
Содержание Multi 3620 IDS
Страница 1: ...OPERATING MANUAL ba77169d06 10 2021 Multi 3620 IDS DIGITAL METER FOR IDS SENSORS 3620 IDS...
Страница 2: ...Multi 3620 IDS 2 ba77169d06 10 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3620 IDS Commissioning ba77169d06 10 2021 15 3 3 4 Setting the date and time 1 See section 4 5 5...
Страница 110: ...Glossary Multi 3620 IDS 110 ba77169d06 10 2021...
Страница 113: ......