Message
Priority
Audible Signal
Visible Signal (LED)
Warning
Low priority
(LP)
User awareness is required,
optimal perfusion compromised
.
E C
— __
Yellow: general alarms
Cyan: temperature related alarms
Error
Medium priority
(MP)
Prompt user response is required,
else fallback to cold storage.
C C C
— — —
Yellow: general alarms
Cyan: temperature related alarms
Only one manufacturer-configured alarm preset is available which is automatically restored
after power interruption.
The generation of the alarm cannot be inactivated. The audible signal can be temporarily
paused by pressing the
‘pause audible alarm button’, this will disable the audible signal for 3
minutes meanwhile the visible alarm signal will remain. After 3 minutes, the audible alarm will
continue.
The alarm signal shall not automatically cease when its triggering event no longer exists; to
reset the alarm, press the OK-button.
The delay for the system to positively identify an alarm state is about 3 seconds.