Turn on the oxygen supply and adjust oxygen flow such that required saturation is
reached (please check the user manual of the oxygenator for maximal flow).
•
Cannulate aortic stump (24 Fr, straight) and portal vein (24 Fr, bend).
•
Open the sterile drape to create a sterile area.
•
Place liver in reservoir.
•
De-air and connect cannulas to corresponding inlet tubing.
•
Start arterial and portal perfusion.
•
Close reservoir with inner and outer lid.
Figure 15: Pressure loss corrections
0
50
100
150
200
250
300
0
200
400
600
800
1000
P
re
ss
ur
e
dr
op
[m
m
H
g]
Flow [ml/min]
8Fr
10Fr
12Fr
0
1
2
3
4
5
0
500
1000
1500
2000
2500
3000
Pr
e
ssur
e
d
ro
p
[
mmH
g]
Flow [ml/min]
25Fr (PV)
25Fr (HA)
When using small cannulas (<24Fr), be aware of the pressure drop. See figure
below for correction.