11.01.601.7c EN
Page 35 of 40
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% UT
for 0,5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% UT
for 1 cycle
70% UT
for 25/30 cycles
0% U
for 250/300 cycles
0% UT
for 0,5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% UT
for 1 cycle
70% UT
for 25/30 cycles
0% U
for 250/300 cycles
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Liver Assist requires continued operation
during power mains interruptions, it is
recommended that the Liver Assist be powered
from an uninterruptible power supply or a
battery.
* Temporary, self-recoverable loss of function is
allowed.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Conducted RF
IEC 61000-4-6
3V
0,15 to 80 MHz
6 V in ISM bands
between 0,15 & 80
MHz
80 % AM at 1 kHz
3V
0,15 to 80 MHz
6 V in ISM bands
between 0,15 & 80
MHz
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter to any part of the device
including cables.
Proximity RF fields
IEC6100-4-3
3 V/m
see table 4
3 V/m
see table 4
Portable RF communications equipment
(including peripherals such as antenna cables
and external antennas) should be used no
closer than 30 cm (12 inches)
to any part of the device, including cables
specified by XVIVO.
NOTE: UT is the a.c. mains voltage prior to application of the test level.