11.01.601.7c EN
Page 15 of 40
•
Power on the Liver Assist by pushing the power button on the control panel.
•
Wait until the self test is completed.
•
Press the OK-button to activate the setup procedure.
•
Unpack a sterile perfusion circuit.
•
Attach the separately packed pressure sensors to the PV and HA pressure lines of
the perfusion circuit. Take care to connect it tightly and to maintain sterility.
•
Place the organ reservoir into the holder of the table, use the red tubing at the arterial
side (HA) and the blue tubing at the portal side (PV).
•
Place the oxygenators on the holders (figure 10).
Figure 10: Connection of oxygenator to the holder
Connect DISPOSABLE
Press to continue
Connect DISPOSABLE
Press to continue
PORTAL VEIN
HEPATIC ARTERY
Before starting the procedure, check the units, trolley, sensors, cables and
connections. Do not use a damaged device.
A sequence of message screens will guide you through set-up and running. Extra
messages will pop-up when actions are required, see section 6.1.
The firmware version number is displayed during start-up. For full version
information, press the OK button during the self-test to view the build number.