background image

XOS  

|

  801377-02  

|

  Rev D  

|

  English

 

 

 

 

 

Page 

2

 of 

22 

 

 

General Notes

: These notes provide 

important information about the equipment’s 

operation and restrictions, including best 

practices to get the best out of your 

instrument.  Always read these notes to 

ensure proper operation. 

 

Radiation Warning

: These warnings apply 

to ionizing radiation hazards 

 

Electrical Shock Warning

: These 

warnings apply to possibility of electrical 

shock 

 

Hot Surface Warning

: These warnings 

apply to hot surfaces 

 

Dangers, Warnings, and Cautions 

Improper use may result in radiation, electrical, chemical, or 

thermal exposure!  

  Follow all local, state, and federal regulations when 

operating this equipment 

  This product must only be used by authorized and 

adequately trained personnel  

  Do not alter the unit in any way 

Before starting, review all of the following warning statements: 

 

 

Radiation Exposure:

 

CAUTION

:  

THIS PRODUCT PRODUCES X-RAYS 

WHEN ENERGIZED DURING NORMAL USE.  

TO BE OPERATED ONLY BY QUALIFIED 

PERSONNEL.  Operating it as described in 

this guide and the full User Manual ensures 

that radiation safeguards work properly to limit 

radiation exposure, keeping it well below the 

levels that are considered safe. Unintended 

exposure can cause serious injury. Do not try 

to disassemble any portion of the analyzer.  

 

Hazardous Sample Handling

: When 

handling hazardous samples, follow proper 

instructions for their storage, disposal, and 

care. Many samples are flammable; keep them 

away from fire.  Wear proper personal 

protective equipment (PPE). See your facility’s 

standard operating procedures for proper PPE 

usage and consult product SDS.  

 

Beryllium Exposure

: The X-Ray detector 

has parts that contain beryllium. In some 

countries, beryllium is a controlled substance, 

subject to regulations concerning the disposal 

of chemical waste. Follow all proper disposal 

procedures when disposing of this substance.  

Beryllium inhalation may cause serious and/or 

chronic lung disease. Beryllium exposure 

poses a potential cancer hazard, based 

primarily on animal testing. In case of any 

damage to the exposed beryllium window, 

contact XOS to return the unit. 

Содержание Petra 4294

Страница 1: ...801377 02 REV D better analysis counts Basic User Manual...

Страница 2: ...uide Used in any manner not specified by XOS Operated in an unsuitable environment Modified repaired or serviced by anyone other than XOS authorized personnel Subject to an accident mishap disaster or...

Страница 3: ...BY QUALIFIED PERSONNEL Operating it as described in this guide and the full User Manual ensures that radiation safeguards work properly to limit radiation exposure keeping it well below the levels th...

Страница 4: ...ww p65warnings ca gov Radiation Safety Statement This Product conforms to all applicable Standards under U S 21 C F R Chapter I Subchapter J and IEC EN 61010 1 Each product is delivered with a radiati...

Страница 5: ...nge Sulfur 2 6 ppm 10 wt Limit of Detection Sulfur 2 6 ppm at 600s Applications Sulfur in hydrocarbons like crude oil diesel gasoline jet fuel and lubricants Product Specifications Method Compliance A...

Страница 6: ...XOS 801377 02 Rev D English Page 5 of 22 Line Drawing Front View Power Button Screen Drip Tray Radiation Warning Light Sample Lid USB Port...

Страница 7: ...XOS 801377 02 Rev D English Page 6 of 22 Line Drawing Back View Ethernet Port Power Cable Connector USB Port Serial Number Label Sample Lid...

Страница 8: ...f 4 x 10mL 1 Power Supply 160W 100 240V AC input 24V DC output Quantity Description 1 XOS USB containing Petra documentation 8 Petra documentation 1 Radiation Safety Test Report 2 Petra Printer Docume...

Страница 9: ...n and can start using the analyzer Setting the Date Time Data files created during measurements are coded with the date and time the measurement was initiated It is critical that the date and time of...

Страница 10: ...g this action and informing the user that the unit will restart itself f After the unit reboots itself notice the change in the date and time Optional Device Setup To optimize usage you can connect th...

Страница 11: ...urement Shutting Down the Instrument To shut down unit 1 Press the power button for 3 seconds or until the screen turns black 2 After about 30 seconds the power button should no longer illuminate mean...

Страница 12: ...he Petra MAX XOS recommends the use of Etnom sample film This film can be reordered through XOS P N 200335 02 Mylar For the best results with the Petra 4294 XOS recommends the use of Mylar sample film...

Страница 13: ...slide with the sample information 7 When the first measurement has successfully completed the unit will report the results a If the unit is still measuring the results will be reported in the History...

Страница 14: ...a Preset 1 From the Scan Info screen select the button in the Preset section or follow the steps below 2 Find and select Settings located on the left side of screen 3 Locate and open Preset Manager 4...

Страница 15: ...o make this Preset the default the preset selected when the analyzer initially starts up if desired g When done select Save Opening Results 1 From the home screen select the History button 2 If it is...

Страница 16: ...scan file s HDX format is proprietary to Petra and is not compatible with any other software program Transfer to USB 1 Navigate to the Settings screen 2 Select the Transfer Data button 3 Insert an US...

Страница 17: ...you want to remove the USB from the unit i If this is true then select OK ii If you want to copy more files select Cancel 7 Now you can remove the USB from the unit Transfer Data through the Network 1...

Страница 18: ...ll be a security pop up The login is User name pi Password H ppy Gr0C33 7 When the Data folder is opened noticed that there will be date folders and a handful of other folders i e AppData Cal Config D...

Страница 19: ...remove it from the sample interface and use cotton swabs with IPA If the basket is damaged replace it with a new basket Do not touch the sample window as this can affect your measurement results Prop...

Страница 20: ...finement Calibration Set PetraMAX only 204715 01 USB powered hub For a more exhaustive list reference the Full User Manual xos com or call XOS Service Replacement Parts Before starting any maintenance...

Страница 21: ...month If there is a spill or contaminants on the film it should be replaced immediately Replacing the Inlet Fan Filter 1 Gently pry the fan filter cover off the fan guard 2 Remove the filter 3 Lay the...

Страница 22: ...til the message disappears which occurs when the x ray tube has cooled below 60 C The message pop up will NOT clear by pressing OK while the tube temperature is close to 65 C The message auto clear wh...

Страница 23: ...rror IOP temp High The IOP is monitoring the source temperature which is too high Blocked vents fan failure or hot ambient temperature 1 Check for 6 inches of clearance around the analyzer 2 Check the...

Страница 24: ...Contact the XOS Service Team Service 1 518 880 1525 service xos com xos com 15 Tech Valley Drive East Greenbush NY 12061...

Страница 25: ......

Отзывы: