XOS
|
801377-02
|
Rev D
|
English
Page
22
of
22
Error Message
Condition
Possible Cause(s)
Solutions
[ERROR[ERROR_IOP[Error IOP
temp High]]]
The IOP is
monitoring the source
temperature, which is
too high
Blocked vents, fan failure,
or hot ambient
temperature
1) Check for 6 inches of clearance around
the analyzer;
2) Check the fan filter; if dirty, clean and
replace.
To view the full list of possible errors refer to the unit User Manual