XOS
|
801377-02
|
Rev D
|
English
Page
20
of
22
2. With the basket in hand, remove and discard the
secondary film
3. Remove any excess adhesive from the Sample Basket
(Isopropyl Alcohol recommended).
4. Remove the adhesive backing from a new sample
basket window film without touching the film.
5. Place the new film over the upside-down basket. Never
touch the Sample Basket film.
6. Turn the basket right-side up.
7. Lower the basket into the measurement chamber. The
basket should have a snug, flush fit in the chamber.
XOS recommends that the Sample Basket film
be changed at least once a month. If there is a
spill or contaminants on the film, it should be
replaced immediately.
Replacing the Inlet Fan Filter
1. Gently pry the fan filter cover off the fan guard.
2. Remove the filter.
3. Lay the new filter flat inside the fan cover.
4. Replace the fan filter cover on the fan guard.
XOS recommends that the fan filter be replaced
at least once a year. If the testing area collects
a lot of dust the filter might need to be changed
more often.
Onsite Service
For issues outside of routine maintenance refer to the full user
manual, contact XOS Technical Service, or contact your local
XOS distributor.
XOS P/N
Common Name
308700-02 Drip Tray
308627-01 Sample Basket
204473-01 Fan Filter
For a more exhaustive list reference the Full User Manual,
xos.com or call XOS Service.