RS-232 Operation
TM-XDOP-01XN
4–17
Configuration
Front Panel
First set the RS-232 parameters:
Next, select RS-232 as the remote control interface.
Press the LCL/RMT button to begin remote operation via the RS-232 interface.
Step # Do This
You Will See
1
ERROR MSGS
2
5 times
REMOTE CONFIG
3
RS-232 Cfg
4
Baud 9600
5
or
Select the baud rate.
Baud #####
6
Flow Ctl None
7
or
Select the flow control option.
Options are
None
= no flow control
Hdwr
= CTS/DTS hardware handshake
XON
= software, XON/XOFF characters used
8
Settings are saved.
MENU
ENTER
ENTER
ENTER
ENTER
Step # Do This
You Will See
1
ERROR MSGS
2
4 times
REMOTE SELECT
3
RS-232
4
RS-232 setting is saved.
MENU
ENTER
ENTER
TM-XDOP-01XN.book Page 17 Monday, July 17, 2006 11:19 AM
Содержание XDC 10-1200
Страница 2: ......
Страница 3: ...XDC 6000 Watt and 12000 Watt Series Digital Programmable Power Supply Operating Manual ...
Страница 8: ...vi ...
Страница 12: ...x ...
Страница 22: ...xx ...
Страница 24: ...xxii ...
Страница 39: ...2 Installation Chapter 2 explains how to mount and wire the XDC and also how to perform basic functional tests ...
Страница 43: ...Location Mounting and Ventilation TM XDOP 01XN 2 5 Figure 2 2 Unpacking the Power Supply 6000 Watt shown ...
Страница 48: ...Installation 2 10 TM XDOP 01XN Figure 2 5 Attaching the AC Input Wires for 6000 Watt units ...
Страница 59: ...Load Connections TM XDOP 01XN 2 21 Figure 2 9 Output for 12000 Watt units low and medium voltage ...
Страница 61: ...Load Connections TM XDOP 01XN 2 23 Figure 2 11 Output for 12000 Watt units high voltage 300 600V ...
Страница 114: ...Operation 3 52 TM XDOP 01XN ...
Страница 208: ...A 16 ...
Страница 224: ...B 16 ...
Страница 259: ...D Error Messages Appendix D lists all the error codes and their meanings ...
Страница 268: ...D 10 ...
Страница 269: ...E GPIB Appendix E describes the General Purpose Interface Bus GPIB commands and lines supported by this model ...
Страница 283: ......