![X-Rite X-Rite Скачать руководство пользователя страница 81](http://html2.mh-extra.com/html/x-rite/x-rite/x-rite_user-manual_3106795081.webp)
X - R i t e e X a c t ™ I n s t r u m e n t
79
Replacing the Battery Pack
CAUTION:
Risk of explosion if battery is replaced by an incorrect type. Replace only with
X-Rite part number SE15-44, self protected rechargeable lithium-ion battery pack.
1.
Release the target base lock and carefully turn the instrument on its side as shown.
2.
Press in on the front edge of the battery access cover (1) while pulling outward to remove.
3.
Disconnect the battery connector (2) by pressing in on the locking tab and unplugging the
connector.
Do not
disconnect by pulling on the battery wires.
4.
Remove the old battery pack (3) from the battery compartment
5.
Insert the new battery pack into the battery compartment with the connector facing
upwards.
6.
Plug the battery into the connector. Press in until the connector is properly seated.
7.
Insert the back edge of the access cover into the bottom of the instrument and rotate the
cover in until it snaps into position.
8.
Reset the battery replacement time. To perform this procedure, tap the Diagnostics tool
icon
on the main menu and select the battery icon
. Tap the New Battery button.
(1)
(1)
(3)
(2)