Configuration
87
Selection of recorder output REC 1
(ST / R / RS models)
Corresponding to the set measuring modes and ranges, the suitable recorder measuring ranges
can be selected.
Measuring
ranges
1
0 ... 199.9
µS/cm
2
0 ... 1999
µS/cm
3
0 ... 19.99 mS/cm
4
0 ... 199.9
mS/cm
Salinity
0.0 ... 70.0
Temperature
-5 ... +60°C
REC ranges
fixed
variable
variable
variable
variable
variable
min. (lower limit of
measuring range)
0
µS/cm
0 µS/cm
0 mS/cm
0 mS/cm
0.0
-5°C... +5°C
Minimum range
∆
of 500 µS/cm,
freely movable
in measuring
range
∆
of 5 mS/cm,
freely movable in
measuring range
∆
of 50 mS/cm,
freely movable in
measuring range
∆
of 10,
freely movable
in measuring
range
+5°C... +30°C
max. (upper limit
of measuring
range)
199.9 µS/cm 1999 µS/cm
19.99 mS/cm
199.9 mS/cm
70.0
+25°C...+60°C
D/span
(step size for
setting)
------
100 µS/cm
1 mS/cm
10 mS/cm
5
1°C
When automatic selection of measuring range (CONF RANGE: AUTO) is set the
REC range is always: 0 ... 199.9
mS/cm (measuring range 4).
Setting:
Submenu "CONF REC 1"
With
enter submenu "CONF REC 1".
The upper conductivity value,
e. g. 1999 µS/cm, flashes.
With
setting in 100 µS steps
between 500 and 1999 µS/cm.
(Minimum range
∆
500 µS/cm).
Set desired upper conductivity value.
With
go on to lower conductivity value.
The lower conductivity value,
e. g. 0 µS/cm, flashes.
With
setting in 100 µS steps
between 0 and 1500 µS/cm.
(Minimum range
∆
500 µS/cm).
Set desired lower conductivity value.
With
go on to "CONF REC 2".