![wtw TecnoLine LF 171 Скачать руководство пользователя страница 101](http://html1.mh-extra.com/html/wtw/tecnoline-lf-171/tecnoline-lf-171_operating-manual_1000707101.webp)
Parameterization
105
Temperature compensation
(ST / R / RS models)
This submenu facilitates to set the temperature function for the temperature compensation. You
can select between a linear temperature coefficient with setting possibility from 0.00 .. 3.00 %/K
and a non-linear function (nLF).
Setting:
Submenu "PARA T-C":
With
enter submenu "PARA T-C".
Indication of measured value and linear
temperature coefficient (example: 1.88).
With
set linear temperature
coefficient between
0.00 and 3.00 %/K.
Below 0.00 and above 3.00 the instrument
automatically switches over to nLF.
With
go on to "PARA REC 1".
With nLF the display shows:
nLF
Operation.