![wtw TecnoLine LF 171 Скачать руководство пользователя страница 109](http://html1.mh-extra.com/html/wtw/tecnoline-lf-171/tecnoline-lf-171_operating-manual_1000707109.webp)
Queries / Settings
113
Conductivity check (Calibration)
For each step a user guidance (indication of the correct operation key) is displayed.
C
heck
After depressing
C
heck
the recorder outputs and the transmitting
relays are frozen.
Timer functions remain unchanged.
or
with
or
M
you can go back to measurement in the
operation level.
The following values are displayed:
⇐
⇐
Indication of the configured
conductivity cell.
⇐
⇐
The value of a standard calibration
solution is indicated in µS or mS.
⇐
⇐
The cell constant can be altered in the
range ± 10 % with the
keys:
Keeping the key depressed means
scrolling up or down, starting with the last
digit.
With pressing
or
M
go back to measuring operation.
During calibration or setting of the cell constant a temperature compensation with reference
temperature 25°C is automatically performed. The temperature coefficient of a 0.01 molar
potassium chloride solution according to standard EN 27 888 has been set fixed.